Filtered by vendor Ge
Subscriptions
Total
128 CVE
CVE | Vendors | Products | Updated | CVSS v3.1 |
---|---|---|---|---|
CVE-2021-27438 | 1 Ge | 2 Reason Dr60, Reason Dr60 Firmware | 2024-11-21 | 8.8 High |
The software contains a hard-coded password it uses for its own inbound authentication or for outbound communication to external components on the Reason DR60 (all firmware versions prior to 02A04.1). | ||||
CVE-2021-27430 | 1 Ge | 1 Ur Bootloader Binary | 2024-11-21 | 8.4 High |
GE UR bootloader binary Version 7.00, 7.01 and 7.02 included unused hardcoded credentials. Additionally, a user with physical access to the UR IED can interrupt the boot sequence by rebooting the UR. | ||||
CVE-2021-27428 | 1 Ge | 38 Multilin B30, Multilin B30 Firmware, Multilin B90 and 35 more | 2024-11-21 | 9.8 Critical |
GE UR IED firmware versions prior to version 8.1x supports upgrading firmware using UR Setup configuration tool – Enervista UR Setup. This UR Setup tool validates the authenticity and integrity of firmware file before uploading the UR IED. An illegitimate user could upgrade firmware without appropriate privileges. The weakness is assessed, and mitigation is implemented in firmware Version 8.10. | ||||
CVE-2021-27426 | 1 Ge | 38 Multilin B30, Multilin B30 Firmware, Multilin B90 and 35 more | 2024-11-21 | 9.8 Critical |
GE UR IED firmware versions prior to version 8.1x with “Basic” security variant does not allow the disabling of the “Factory Mode,” which is used for servicing the IED by a “Factory” user. | ||||
CVE-2021-27424 | 1 Ge | 38 Multilin B30, Multilin B30 Firmware, Multilin B90 and 35 more | 2024-11-21 | 5.3 Medium |
GE UR firmware versions prior to version 8.1x shares MODBUS memory map as part of the communications guide. GE was made aware a “Last-key pressed” MODBUS register can be used to gain unauthorized information. | ||||
CVE-2021-27422 | 1 Ge | 38 Multilin B30, Multilin B30 Firmware, Multilin B90 and 35 more | 2024-11-21 | 7.5 High |
GE UR firmware versions prior to version 8.1x web server interface is supported on UR over HTTP protocol. It allows sensitive information exposure without authentication. | ||||
CVE-2021-27420 | 1 Ge | 38 Multilin B30, Multilin B30 Firmware, Multilin B90 and 35 more | 2024-11-21 | 5.3 Medium |
GE UR firmware versions prior to version 8.1x web server task does not properly handle receipt of unsupported HTTP verbs, resulting in the web server becoming temporarily unresponsive after receiving a series of unsupported HTTP requests. When unresponsive, the web server is inaccessible. By itself, this is not particularly significant as the relay remains effective in all other functionality and communication channels. | ||||
CVE-2021-27418 | 1 Ge | 38 Multilin B30, Multilin B30 Firmware, Multilin B90 and 35 more | 2024-11-21 | 5.3 Medium |
GE UR firmware versions prior to version 8.1x supports web interface with read-only access. The device fails to properly validate user input, making it possible to perform cross-site scripting attacks, which may be used to send a malicious script. Also, UR Firmware web server does not perform HTML encoding of user-supplied strings. | ||||
CVE-2020-6992 | 1 Ge | 1 Cimplicity | 2024-11-21 | 6.7 Medium |
A local privilege escalation vulnerability has been identified in the GE Digital CIMPLICITY HMI/SCADA product v10.0 and prior. If exploited, this vulnerability could allow an adversary to modify the system, leading to the arbitrary execution of code. This vulnerability is only exploitable if an attacker has access to an authenticated session. GE Digital CIMPLICITY v11.0, released January 2020, contains mitigation for this local privilege escalation vulnerability. GE Digital recommends all users upgrade to GE CIMPLICITY v11.0 or newer. | ||||
CVE-2020-6977 | 1 Ge | 32 Invenia Abus Scan Station, Invenia Abus Scan Station Firmware, Logiq E10 and 29 more | 2024-11-21 | 6.8 Medium |
A restricted desktop environment escape vulnerability exists in the Kiosk Mode functionality of affected devices. Specially crafted inputs can allow the user to escape the restricted environment, resulting in access to the underlying operating system. Affected devices include the following GE Ultrasound Products: Vivid products - all versions; LOGIQ - all versions not including LOGIQ 100 Pro; Voluson - all versions; Versana Essential - all versions; Invenia ABUS Scan station - all versions; Venue - all versions not including Venue 40 R1-3 and Venue 50 R4-5 | ||||
CVE-2020-36549 | 1 Ge | 2 Voluson S8, Voluson S8 Firmware | 2024-11-21 | 8.8 High |
A vulnerability classified as critical was found in GE Voluson S8. Affected is the underlying Windows XP operating system. Missing patches might introduce an excessive attack surface. Access to the local network is required for this attack to succeed. | ||||
CVE-2020-36548 | 1 Ge | 2 Voluson S8, Voluson S8 Firmware | 2024-11-21 | 5.9 Medium |
A vulnerability classified as problematic has been found in GE Voluson S8. Affected is the file /uscgi-bin/users.cgi of the Service Browser. The manipulation leads to improper authentication and elevated access possibilities. It is possible to launch the attack on the local host. | ||||
CVE-2020-36547 | 1 Ge | 2 Voluson S8, Voluson S8 Firmware | 2024-11-21 | 5.9 Medium |
A vulnerability was found in GE Voluson S8. It has been rated as critical. This issue affects the Service Browser which itroduces hard-coded credentials. Attacking locally is a requirement. It is recommended to change the configuration settings. | ||||
CVE-2020-27267 | 4 Ge, Ptc, Rockwellautomation and 1 more | 7 Industrial Gateway Server, Kepware Kepserverex, Opc-aggregator and 4 more | 2024-11-21 | 9.1 Critical |
KEPServerEX v6.0 to v6.9, ThingWorx Kepware Server v6.8 and v6.9, ThingWorx Industrial Connectivity (all versions), OPC-Aggregator (all versions), Rockwell Automation KEPServer Enterprise, GE Digital Industrial Gateway Server v7.68.804 and v7.66, and Software Toolbox TOP Server all 6.x versions, are vulnerable to a heap-based buffer overflow. Opening a specifically crafted OPC UA message could allow an attacker to crash the server and potentially leak data. | ||||
CVE-2020-27265 | 4 Ge, Ptc, Rockwellautomation and 1 more | 7 Industrial Gateway Server, Kepware Kepserverex, Opc-aggregator and 4 more | 2024-11-21 | 9.8 Critical |
KEPServerEX: v6.0 to v6.9, ThingWorx Kepware Server: v6.8 and v6.9, ThingWorx Industrial Connectivity: All versions, OPC-Aggregator: All versions, Rockwell Automation KEPServer Enterprise, GE Digital Industrial Gateway Server: v7.68.804 and v7.66, Software Toolbox TOP Server: All 6.x versions are vulnerable to a stack-based buffer overflow. Opening a specifically crafted OPC UA message could allow an attacker to crash the server and remotely execute code. | ||||
CVE-2020-27263 | 4 Ge, Ptc, Rockwellautomation and 1 more | 7 Industrial Gateway Server, Kepware Kepserverex, Opc-aggregator and 4 more | 2024-11-21 | 9.1 Critical |
KEPServerEX: v6.0 to v6.9, ThingWorx Kepware Server: v6.8 and v6.9, ThingWorx Industrial Connectivity: All versions, OPC-Aggregator: All versions, Rockwell Automation KEPServer Enterprise, GE Digital Industrial Gateway Server: v7.68.804 and v7.66, Software Toolbox TOP Server: All 6.x versions, are vulnerable to a heap-based buffer overflow. Opening a specifically crafted OPC UA message could allow an attacker to crash the server and potentially leak data. | ||||
CVE-2020-25197 | 1 Ge | 6 Rt430, Rt430 Firmware, Rt431 and 3 more | 2024-11-21 | 9.8 Critical |
A code injection vulnerability exists in one of the webpages in GE Reason RT430, RT431 & RT434 GNSS clocks in firmware versions prior to version 08A06 that could allow an authenticated remote attacker to execute arbitrary code on the system. | ||||
CVE-2020-25193 | 1 Ge | 6 Rt430, Rt430 Firmware, Rt431 and 3 more | 2024-11-21 | 5.3 Medium |
By having access to the hard-coded cryptographic key for GE Reason RT430, RT431 & RT434 GNSS clocks in firmware versions prior to version 08A06, attackers would be able to intercept and decrypt encrypted traffic through an HTTPS connection. | ||||
CVE-2020-16246 | 1 Ge | 4 S2020, S2020 Firmware, S2024 and 1 more | 2024-11-21 | 6.1 Medium |
The affected Reason S20 Ethernet Switch is vulnerable to cross-site scripting (XSS), which may allow attackers to trick users into following a link or navigating to a page that posts a malicious JavaScript statement to the vulnerable site, causing the malicious JavaScript to be rendered by the site and executed by the victim client. | ||||
CVE-2020-16244 | 1 Ge | 1 Asset Performance Management Classic | 2024-11-21 | 7.2 High |
GE Digital APM Classic, Versions 4.4 and prior. Salt is not used for hash calculation of passwords, making it possible to decrypt passwords. This design flaw, along with the IDOR vulnerability, puts the entire platform at high risk because an authenticated user can retrieve all user account data and then retrieve the actual passwords. |