Filtered by CWE-259
Total 100 CVE
CVE Vendors Products Updated CVSS v3.1
CVE-2014-5431 1 Baxter 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module 2024-08-06 N/A
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.
CVE-2014-5434 1 Baxter 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module 2024-08-06 N/A
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
CVE-2015-3953 1 Pifzer 6 Plum A\+3 Infusion System, Plum A\+3 Infusion System Firmware, Plum A\+ Infusion System and 3 more 2024-08-06 N/A
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
CVE-2016-9358 1 Marel 44 A320, A320 Firmware, A325 and 41 more 2024-08-06 N/A
A Hard-Coded Passwords issue was discovered in Marel Food Processing Systems M3000 terminal associated with the following systems: A320, A325, A371, A520 Master, A520 Slave, A530, A542, A571, Check Bin Grader, FlowlineQC T376, IPM3 Dual Cam v132, IPM3 Dual Cam v139, IPM3 Single Cam v132, P520, P574, SensorX13 QC flow line, SensorX23 QC Master, SensorX23 QC Slave, Speed Batcher, T374, T377, V36, V36B, and V36C; M3210 terminal associated with the same systems as the M3000 terminal identified above; M3000 desktop software associated with the same systems as the M3000 terminal identified above; MAC4 controller associated with the same systems as the M3000 terminal identified above; SensorX23 X-ray machine; SensorX25 X-ray machine; and MWS2 weighing system. The end user does not have the ability to change system passwords.
CVE-2017-20039 1 Sicunet 1 Access Control 2024-08-05 9.8 Critical
A vulnerability was found in SICUNET Access Controller 0.32-05z. It has been classified as very critical. This affects an unknown part. The manipulation leads to weak authentication. It is possible to initiate the attack remotely.
CVE-2017-6039 1 Phoenixbroadband 2 Poweragent Sc3 Bms, Poweragent Sc3 Bms Firmware 2024-08-05 N/A
A Use of Hard-Coded Password issue was discovered in Phoenix Broadband PowerAgent SC3 BMS, all versions prior to v6.87. Use of a hard-coded password may allow unauthorized access to the device.
CVE-2017-6022 1 Bd 2 Kla Journal Service, Performa 2024-08-05 N/A
A hard-coded password issue was discovered in Becton, Dickinson and Company (BD) PerformA, Version 2.0.14.0 and prior versions, and KLA Journal Service, Version 1.0.51 and prior versions. They use hard-coded passwords to access the BD Kiestra Database, which could be leveraged to compromise the confidentiality of limited PHI/PII information stored in the BD Kiestra Database.
CVE-2024-38885 1 Horizoncloud 1 Caterease 2024-08-05 7.5 High
An issue in Horizon Business Services Inc. Caterease 16.0.1.1663 through 24.0.1.2405 and possibly later versions, allows a remote attacker to perform unauthorized access using known operating system credentials due to hardcoded SQL user credentials in the client application.
CVE-2018-25069 1 Netis-systems 2 Netcore Router, Netcore Router Firmware 2024-08-05 7.3 High
A vulnerability classified as critical has been found in Netis Netcore Router. This affects an unknown part. The manipulation leads to use of hard-coded password. It is possible to initiate the attack remotely. The identifier VDB-217593 was assigned to this vulnerability.
CVE-2019-13530 1 Philips 19 865240, 865241, 865242 and 16 more 2024-08-04 7.2 High
Philips IntelliVue WLAN, portable patient monitors, WLAN Version A, Firmware A.03.09, WLAN Version A, Firmware A.03.09, Part #: M8096-67501, WLAN Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C) and WLAN Version B, Firmware A.01.09, Part #: N/A (Replaced by Version C). An attacker can use these credentials to login via ftp and upload a malicious firmware.
CVE-2019-10881 1 Xerox 20 Altalink B8045, Altalink B8045 Firmware, Altalink B8055 and 17 more 2024-08-04 9.8 Critical
Xerox AltaLink B8045/B8055/B8065/B8075/B8090, AltaLink C8030/C8035/C8045/C8055/C8070 with software releases before 103.xxx.030.32000 includes two accounts with weak hard-coded passwords which can be exploited and allow unauthorized access which cannot be disabled.
CVE-2019-3908 1 Identicard 1 Premisys Id 2024-08-04 7.5 High
Premisys Identicard version 3.1.190 stores backup files as encrypted zip files. The password to the zip is hard-coded and unchangeable. An attacker with access to these backups can decrypt them and obtain sensitive data.
CVE-2020-12039 1 Baxter 2 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware 2024-08-04 2.4 Low
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.
CVE-2020-12045 1 Baxter 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module 2024-08-04 9.8 Critical
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials.
CVE-2020-12016 1 Baxter 4 Em1200, Em1200 Firmware, Em2400 and 1 more 2024-08-04 9.8 Critical
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account credentials for the ExactaMix operating system. Successful exploitation of this vulnerability may allow an attacker who has gained unauthorized access to system resources, including access to execute software or to view/update files, directories, or system configuration. This could allow an attacker with network access to view sensitive data including PHI.
CVE-2020-12047 1 Baxter 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module 2024-08-04 9.8 Critical
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded credentials.
CVE-2020-12012 1 Baxter 4 Em1200, Em1200 Firmware, Em2400 and 1 more 2024-08-04 6.1 Medium
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. Successful exploitation of this vulnerability may allow an attacker with physical access to gain unauthorized access to view/update system configuration or data. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI.
CVE-2020-12037 1 Baxter 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more 2024-08-04 7.5 High
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
CVE-2020-7590 1 Siemens 2 Dca Vantage Analyzer, Dca Vantage Analyzer Firmware 2024-08-04 6.8 Medium
A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.
CVE-2021-27452 1 Ge 2 Mu320e, Mu320e Firmware 2024-08-03 7.8 High
The software contains a hard-coded password that could allow an attacker to take control of the merging unit using these hard-coded credentials on the MU320E (all firmware versions prior to v04A00.1).