Filtered by vendor Medtronicdiabetes
Subscriptions
Filtered by product 522 Paradigm Real-time
Subscriptions
Total
1 CVE
CVE | Vendors | Products | Updated | CVSS v3.1 |
---|---|---|---|---|
CVE-2018-14781 | 1 Medtronicdiabetes | 18 508 Minimed Insulin Pump, 508 Minimed Insulin Pump Firmware, 522 Paradigm Real-time and 15 more | 2024-09-17 | N/A |
Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the "easy bolus" and "remote bolus" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery. |
Page 1 of 1.