Filtered by vendor Baxter
Subscriptions
Filtered by product Life2000 Ventilator Firmware
Subscriptions
Total
4 CVE
CVE | Vendors | Products | Updated | CVSS v3.1 |
---|---|---|---|---|
CVE-2024-48971 | 1 Baxter | 1 Life2000 Ventilator Firmware | 2024-11-15 | 9.3 Critical |
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges. | ||||
CVE-2024-48966 | 1 Baxter | 1 Life2000 Ventilator Firmware | 2024-11-15 | 10 Critical |
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance. | ||||
CVE-2024-9834 | 1 Baxter | 1 Life2000 Ventilator Firmware | 2024-11-15 | 9.3 Critical |
Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance. | ||||
CVE-2024-48967 | 1 Baxter | 1 Life2000 Ventilator Firmware | 2024-11-15 | 10 Critical |
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance. |
Page 1 of 1.