Filtered by vendor Medtronic
Subscriptions
Filtered by product Mycarelink 24952 Patient Monitor
Subscriptions
Total
2 CVE
CVE | Vendors | Products | Updated | CVSS v3.1 |
---|---|---|---|---|
CVE-2018-10626 | 1 Medtronic | 4 Mycarelink 24950 Patient Monitor, Mycarelink 24950 Patient Monitor Firmware, Mycarelink 24952 Patient Monitor and 1 more | 2024-11-21 | N/A |
A vulnerability was discovered in all versions of Medtronic MyCareLink 24950 and 24952 Patient Monitor. The affected product's update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network. | ||||
CVE-2018-10622 | 1 Medtronic | 4 Mycarelink 24950 Patient Monitor, Mycarelink 24950 Patient Monitor Firmware, Mycarelink 24952 Patient Monitor and 1 more | 2024-11-21 | N/A |
A vulnerability was discovered in all versions of Medtronic MyCareLink 24950 and 24952 Patient Monitor. The affected products use per-product credentials that are stored in a recoverable format. An attacker can use these credentials for network authentication and encryption of local data at rest. |
Page 1 of 1.