Filtered by vendor Baxter
Subscriptions
Filtered by product Prismax
Subscriptions
Total
3 CVE
CVE | Vendors | Products | Updated | CVSS v3.1 |
---|---|---|---|---|
CVE-2020-12037 | 1 Baxter | 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more | 2024-11-21 | 7.5 High |
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device. | ||||
CVE-2020-12036 | 1 Baxter | 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more | 2024-11-21 | 7.5 High |
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device. | ||||
CVE-2020-12035 | 1 Baxter | 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more | 2024-11-21 | 4.9 Medium |
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration. |
Page 1 of 1.