Filtered by vendor Baxter Subscriptions
Filtered by product Prismax Firmware Subscriptions
Total 3 CVE
CVE Vendors Products Updated CVSS v3.1
CVE-2020-12037 1 Baxter 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more 2024-11-21 7.5 High
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
CVE-2020-12036 1 Baxter 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more 2024-11-21 7.5 High
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
CVE-2020-12035 1 Baxter 4 Prismaflex, Prismaflex Firmware, Prismax and 1 more 2024-11-21 4.9 Medium
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration.