Metrics
Affected Vendors & Products
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2014-5289 | Hospira MedNet software version 5.8 and prior uses vulnerable versions of the JBoss Enterprise Application Platform software that may allow unauthenticated users to execute arbitrary code on the target system. Hospira has developed a new version of the MedNet software, MedNet 6.1. Existing versions of MedNet can be upgraded to MedNet 6.1. |
Solution
Hospira has developed a new version of the MedNet software, MedNet 6.1. Hospira reports that MedNet 6.1 no longer uses hard-coded passwords, hard-coded cryptographic keys, and no longer stores passwords in clear text. Existing versions of MedNet can be upgraded to MedNet 6.1. Hospira has produced mitigation recommendations that help mitigate the vulnerability in the vulnerable version of JBoss Enterprise Application Platform software, used in the MedNet software. This has been addressed by Hospira through issuance of the following knowledge based articles: Improving Security in Hospira MedNet 5.5 (August 2014) and Improving Security in Hospira MedNet 5.8 (August 2014). For additional information about Hospira’s new releases and mitigation recommendations, contact Hospira’s technical support at 1-800-241-4002.
Workaround
No workaround given by the vendor.
Mon, 03 Nov 2025 18:45:00 +0000
| Type | Values Removed | Values Added |
|---|---|---|
| Title | Hospira MedNet Code Injection | |
| References |
|
Projects
Sign in to view the affected projects.
Status: PUBLISHED
Assigner: icscert
Published:
Updated: 2025-11-03T18:20:04.649Z
Reserved: 2014-08-22T00:00:00
Link: CVE-2014-5401
No data.
Status : Modified
Published: 2019-03-26T17:29:00.340
Modified: 2025-11-03T19:15:38.057
Link: CVE-2014-5401