Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Advisories
Source ID Title
EUVD EUVD EUVD-2015-3985 Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Fixes

Solution

No solution given by the vendor.


Workaround

No workaround given by the vendor.

History

No history.

cve-icon MITRE

Status: PUBLISHED

Assigner: icscert

Published:

Updated: 2024-08-06T06:04:01.128Z

Reserved: 2015-05-12T00:00:00

Link: CVE-2015-3954

cve-icon Vulnrichment

No data.

cve-icon NVD

Status : Modified

Published: 2019-03-25T17:29:00.670

Modified: 2024-11-21T02:30:08.203

Link: CVE-2015-3954

cve-icon Redhat

No data.

cve-icon OpenCVE Enrichment

No data.