Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
History

No history.

cve-icon MITRE

Status: PUBLISHED

Assigner: icscert

Published: 2018-04-25T13:00:00Z

Updated: 2024-09-17T01:55:36.387Z

Reserved: 2017-08-09T00:00:00

Link: CVE-2017-12716

cve-icon Vulnrichment

No data.

cve-icon NVD

Status : Modified

Published: 2018-04-25T13:29:00.333

Modified: 2024-11-21T03:10:05.107

Link: CVE-2017-12716

cve-icon Redhat

No data.