Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
History

No history.

cve-icon MITRE

Status: PUBLISHED

Assigner: icscert

Published: 2020-06-29T13:45:58

Updated: 2024-08-04T11:48:57.530Z

Reserved: 2020-04-21T00:00:00

Link: CVE-2020-12040

cve-icon Vulnrichment

No data.

cve-icon NVD

Status : Analyzed

Published: 2020-06-29T14:15:11.677

Modified: 2020-07-09T17:58:57.613

Link: CVE-2020-12040

cve-icon Redhat

No data.