A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When HMI3 starts up, it binds a local service to a TCP port on all interfaces of the device, and takes extensive time for the GUI to connect to the TCP socket, allowing the connection to be hijacked by an external attacker.
History

No history.

cve-icon MITRE

Status: PUBLISHED

Assigner: mitre

Published: 2021-08-02T12:44:43

Updated: 2024-08-04T01:16:02.369Z

Reserved: 2021-07-21T00:00:00

Link: CVE-2021-37166

cve-icon Vulnrichment

No data.

cve-icon NVD

Status : Modified

Published: 2021-08-02T13:15:07.897

Modified: 2024-11-21T06:14:46.323

Link: CVE-2021-37166

cve-icon Redhat

No data.