An improper access control vulnerability may allow privilege escalation.This issue affects: 

* ELI 380 Resting Electrocardiograph:

Versions 2.6.0 and prior; 
* ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph:

Versions 2.3.1 and prior; 
* ELI 250c/BUR 250c Resting Electrocardiograph: Versions 2.1.2 and prior; 
* ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph:

Versions 2.2.0 and prior.
Fixes

Solution

Hillrom has released software updates for all impacted devices to address these vulnerabilities. New product versions that mitigate these vulnerabilities are available as follows: * Welch Allyn ELI 380 Resting Electrocardiograph: available Q4 2023 * Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: available May 2022 * Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: available Q4 2023 Hillrom recommends users upgrade to the latest product versions. Information on how to update these products can be found on the Hillrom disclosure page https://hillrom.com/en/responsible-disclosures/ .


Workaround

Hillrom recommends the following workarounds to help reduce risk: * Apply proper network and physical security controls. * Ensure a unique encryption key is configured for ELI Link and Cardiograph. * Where possible, use a firewall to prevent communication on Port 21 FTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet service.

History

Fri, 07 Feb 2025 19:15:00 +0000

Type Values Removed Values Added
Metrics ssvc

{'options': {'Automatable': 'no', 'Exploitation': 'none', 'Technical Impact': 'partial'}, 'version': '2.0.3'}


Fri, 07 Feb 2025 17:15:00 +0000

Type Values Removed Values Added
Description An improper access control vulnerability may allow privilege escalation.This issue affects:  * ELI 380 Resting Electrocardiograph: Versions 2.6.0 and prior;  * ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: Versions 2.3.1 and prior;  * ELI 250c/BUR 250c Resting Electrocardiograph: Versions 2.1.2 and prior;  * ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: Versions 2.2.0 and prior.
Title Improper Access Control Vulnerability in ELI Electrocardiograph Devices
Weaknesses CWE-284
References
Metrics cvssV3_1

{'score': 7.7, 'vector': 'CVSS:3.1/AV:N/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:H'}


cve-icon MITRE

Status: PUBLISHED

Assigner: Baxter

Published:

Updated: 2025-02-07T18:49:22.506Z

Reserved: 2022-03-03T22:04:24.534Z

Link: CVE-2022-26389

cve-icon Vulnrichment

Updated: 2025-02-07T18:48:32.460Z

cve-icon NVD

Status : Received

Published: 2025-02-07T17:15:22.130

Modified: 2025-02-07T17:15:22.130

Link: CVE-2022-26389

cve-icon Redhat

No data.

cve-icon OpenCVE Enrichment

No data.