Description
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Published: 2024-11-14
Score: 9.3 Critical
EPSS: < 1% Very Low
KEV: No
Impact: n/a
Action: n/a
AI Analysis

No analysis available yet.

Remediation

No remediation available yet.

Tracking

Sign in to view the affected projects.

Advisories
Source ID Title
EUVD EUVD EUVD-2024-50148 There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
History

Mon, 18 Nov 2024 16:15:00 +0000

Type Values Removed Values Added
First Time appeared Baxter
Baxter life2000 Ventilator Firmware
CPEs cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*
Vendors & Products Baxter
Baxter life2000 Ventilator Firmware
Metrics ssvc

{'options': {'Automatable': 'no', 'Exploitation': 'none', 'Technical Impact': 'total'}, 'version': '2.0.3'}


Thu, 14 Nov 2024 21:15:00 +0000

Type Values Removed Values Added
Description There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Title No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator
Weaknesses CWE-307
References
Metrics cvssV3_1

{'score': 9.3, 'vector': 'CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H'}


Subscriptions

Baxter Life2000 Ventilator Firmware
cve-icon MITRE

Status: PUBLISHED

Assigner: Baxter

Published:

Updated: 2024-11-18T15:37:00.311Z

Reserved: 2024-10-10T19:24:46.919Z

Link: CVE-2024-9832

cve-icon Vulnrichment

Updated: 2024-11-18T15:36:55.272Z

cve-icon NVD

Status : Awaiting Analysis

Published: 2024-11-14T21:15:22.593

Modified: 2024-11-15T13:58:08.913

Link: CVE-2024-9832

cve-icon Redhat

No data.

cve-icon OpenCVE Enrichment

No data.

Weaknesses