AES-128 encryption key, which is the same across all installations.
Metrics
Affected Vendors & Products
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2025-6389 | A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations. |
Solution
Philips recommends the following mitigations: * Resolved in ISCV 4.2 build 20589, which was released in May 2019. * Philips recommends users upgrade ISCV installed base to the latest ISCV version (at the time of this publication is 830089 – IntelliSpace Cardiovacular 8.0.0.0) * Please contact a local Philips sales (service) representative to learn how to engage this upgrade process. * For managed services users, new releases will be made available upon resource availability. Releases are subject to country-specific regulations. Refer to the Philips advisory https://www.philips.com/a-w/security/security-advisories.html for more details.
Workaround
No workaround given by the vendor.
Sun, 13 Jul 2025 13:45:00 +0000
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
epss
|
epss
|
Thu, 13 Mar 2025 20:15:00 +0000
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
ssvc
|
Thu, 13 Mar 2025 18:30:00 +0000
| Type | Values Removed | Values Added |
|---|---|---|
| Description | A token is created using the username, current date/time, and a fixed AES-128 encryption key, which is the same across all installations. | |
| Title | Philips Intellispace Cardiovascular (ISCV) Use of Weak Credentials | |
| Weaknesses | CWE-1391 | |
| References |
| |
| Metrics |
cvssV3_1
|
Status: PUBLISHED
Assigner: icscert
Published:
Updated: 2025-03-17T20:04:56.100Z
Reserved: 2025-03-11T20:11:49.638Z
Link: CVE-2025-2229
Updated: 2025-03-13T19:30:35.846Z
Status : Received
Published: 2025-03-13T19:15:52.523
Modified: 2025-03-13T19:15:52.523
Link: CVE-2025-2229
No data.
OpenCVE Enrichment
No data.
EUVD