NMIS/BioDose V22.02 and previous version installations where the embedded Microsoft SQLServer Express is used are exposed in the Windows share accessed by clients in networked installs. By default, this directory has insecure directory paths that allow access to the SQL Server database and configuration files, which can contain sensitive data.
Advisories

No advisories yet.

Fixes

Solution

Mirion Medical recommends users update to V23.0 or later. Users with an active support contract should update to the latest version through the software or users can contact Mirion Medical support directly.


Workaround

No workaround given by the vendor.

History

Tue, 02 Dec 2025 22:15:00 +0000

Type Values Removed Values Added
Metrics ssvc

{'options': {'Automatable': 'no', 'Exploitation': 'none', 'Technical Impact': 'partial'}, 'version': '2.0.3'}


Tue, 02 Dec 2025 21:15:00 +0000

Type Values Removed Values Added
Description NMIS/BioDose V22.02 and previous version installations where the embedded Microsoft SQLServer Express is used are exposed in the Windows share accessed by clients in networked installs. By default, this directory has insecure directory paths that allow access to the SQL Server database and configuration files, which can contain sensitive data.
Title Mirion Medical EC2 Software NMIS BioDose Incorrect Permission Assignment for Critical Resource
Weaknesses CWE-732
References
Metrics cvssV3_1

{'score': 8.4, 'vector': 'CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H'}

cvssV4_0

{'score': 8.6, 'vector': 'CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N'}


cve-icon MITRE

Status: PUBLISHED

Assigner: icscert

Published:

Updated: 2025-12-02T21:40:09.463Z

Reserved: 2025-11-11T20:56:52.837Z

Link: CVE-2025-64298

cve-icon Vulnrichment

Updated: 2025-12-02T21:40:05.697Z

cve-icon NVD

Status : Received

Published: 2025-12-02T21:15:52.333

Modified: 2025-12-02T21:15:52.333

Link: CVE-2025-64298

cve-icon Redhat

No data.

cve-icon OpenCVE Enrichment

No data.