Search Results (35 CVEs found)

CVE Vendors Products Updated CVSS v3.1
CVE-2024-48970 1 Baxter 1 Life2000 Ventilator Firmware 2026-04-15 9.3 Critical
The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure.
CVE-2024-9832 1 Baxter 1 Life2000 Ventilator Firmware 2026-04-15 9.3 Critical
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
CVE-2024-9834 1 Baxter 1 Life2000 Ventilator Firmware 2026-04-15 9.3 Critical
Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
CVE-2024-48974 1 Baxter 1 Life2000 Ventilator Firmware 2026-04-15 9.3 Critical
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
CVE-2024-48966 1 Baxter 1 Life2000 Ventilator Firmware 2026-04-15 10 Critical
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
CVE-2024-48967 1 Baxter 1 Life2000 Ventilator Firmware 2026-04-15 10 Critical
The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.
CVE-2024-48973 1 Baxter 1 Life2000 Ventilator Firmware 2026-04-15 9.3 Critical
The debug port on the ventilator's serial interface is enabled by default. This could allow an attacker to send and receive messages over the debug port (which are unencrypted; see 3.2.1) that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
CVE-2024-48971 1 Baxter 1 Life2000 Ventilator Firmware 2026-04-15 9.3 Critical
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.
CVE-2022-26394 1 Baxter 8 Baxter Spectrum Iq 35700bax3, Baxter Spectrum Iq 35700bax3 Firmware, Sigma Spectrum 35700bax and 5 more 2024-11-21 5.5 Medium
The Baxter Spectrum WBM does not perform mutual authentication with the gateway server host. This may allow an attacker to perform a man in the middle attack that modifies parameters making the network connection fail.
CVE-2022-26393 1 Baxter 8 Baxter Spectrum Iq 35700bax3, Baxter Spectrum Iq 35700bax3 Firmware, Sigma Spectrum 35700bax and 5 more 2024-11-21 5 Medium
The Baxter Spectrum WBM is susceptible to format string attacks via application messaging. An attacker could use this to read memory in the WBM to access sensitive information or cause a Denial of Service (DoS) on the WBM.
CVE-2022-26392 1 Baxter 8 Baxter Spectrum Iq 35700bax3, Baxter Spectrum Iq 35700bax3 Firmware, Sigma Spectrum 35700bax and 5 more 2024-11-21 3.1 Low
The Baxter Spectrum WBM (v16, v16D38) and Baxter Spectrum WBM (v17, v17D19, v20D29 to v20D32) when in superuser mode is susceptible to format string attacks via application messaging. An attacker could use this to read memory in the WBM to access sensitive information.
CVE-2022-26390 1 Baxter 8 Baxter Spectrum Iq 35700bax3, Baxter Spectrum Iq 35700bax3 Firmware, Sigma Spectrum 35700bax and 5 more 2024-11-21 4.2 Medium
The Baxter Spectrum Wireless Battery Module (WBM) stores network credentials and PHI (only applicable to Spectrum IQ pumps using auto programming) in unencrypted form. An attacker with physical access to a device that hasn't had all data and settings erased may be able to extract sensitive information.
CVE-2021-43935 1 Baxter 10 Welch Allyn Connex Cardio, Welch Allyn Diagnostic Cardiology Suite, Welch Allyn Hscribe Holter Analysis System and 7 more 2024-11-21 8.1 High
The impacted products, when configured to use SSO, are affected by an improper authentication vulnerability. This vulnerability allows the application to accept manual entry of any active directory (AD) account provisioned in the application without supplying a password, resulting in access to the application as the supplied AD account, with all associated privileges.
CVE-2020-12048 1 Baxter 2 Phoenix X36, Phoenix X36 Firmware 2024-11-21 7.5 High
Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.
CVE-2020-12047 1 Baxter 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module 2024-11-21 9.8 Critical
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded credentials.
CVE-2020-12045 1 Baxter 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module 2024-11-21 9.8 Critical
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials.
CVE-2020-12043 1 Baxter 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module 2024-11-21 9.8 Critical
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operational until the WBM is rebooted.
CVE-2020-12041 1 Baxter 3 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware, Wireless Battery Module 2024-11-21 9.4 Critical
The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to network settings are removed upon reboot.
CVE-2020-12040 1 Baxter 2 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware 2024-11-21 9.8 Critical
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.
CVE-2020-12039 1 Baxter 2 Sigma Spectrum Infusion System, Sigma Spectrum Infusion System Firmware 2024-11-21 2.4 Low
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.