CVE-2020-15483 1 Niscomed 2 M1000 Multipara Patient Monitor, M1000 Multipara Patient Monitor Firmware 2024-08-04 6.8 Medium
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The physical UART debug port provides a shell, without requiring a password, with complete access.
CVE-2020-15484 1 Niscomed 2 M1000 Multipara Patient Monitor, M1000 Multipara Patient Monitor Firmware 2024-08-04 7.5 High
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The internal storage of the underlying Linux system stores data in cleartext, without integrity protection against tampering.
CVE-2020-15485 1 Niscomed 2 M1000 Multipara Patient Monitor, M1000 Multipara Patient Monitor Firmware 2024-08-04 5.5 Medium
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The onboard Flash memory stores data in cleartext, without integrity protection against tampering.
CVE-2020-15482 1 Niscomed 2 M1000 Multipara Patient Monitor, M1000 Multipara Patient Monitor Firmware 2024-08-04 7.8 High
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The device enables an unencrypted TELNET service by default, with a blank password for the admin account. This allows an attacker to gain root access to the device over the local network.