A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections.
Fixes

Solution

No solution given by the vendor.


Workaround

No workaround given by the vendor.

History

No history.

cve-icon MITRE

Status: PUBLISHED

Assigner: icscert

Published:

Updated: 2024-08-05T18:43:56.473Z

Reserved: 2017-08-09T00:00:00

Link: CVE-2017-12724

cve-icon Vulnrichment

No data.

cve-icon NVD

Status : Modified

Published: 2018-02-15T10:29:00.477

Modified: 2024-11-21T03:10:06.060

Link: CVE-2017-12724

cve-icon Redhat

No data.

cve-icon OpenCVE Enrichment

No data.