EUVD-2020-4339 | Philips IntelliBridge Enterprise (IBE), Versions B.12 and prior, IntelliBridge Enterprise system integration with SureSigns (VS4), EarlyVue (VS30) and IntelliVue Guardian (IGS). Unencrypted user credentials received in the IntelliBridge Enterprise (IBE) are logged within the transaction logs, which are secured behind the login based administrative web portal. The unencrypted user credentials sent from the affected products listed above, for the purpose of handshake or authentication with the Enterprise Systems, are logged as the payload in IntelliBridge Enterprise (IBE) within the transaction logs. An attacker with administrative privileges could exploit this vulnerability to read plain text credentials from log files. |
Solution
Philips plans a new release (IBE B.13) by the end of Q4 2020 that remediates the security risk by not logging the plain text user credentials in the log file. Users with questions regarding their specific Philips IntelliBridge Enterprise installation should contact their local Philips service support team, or regional service support. Users can see the Philips advisory http://www.philips.com/productsecurity for more details, and contact the Philips Customer Service Solutions https://www.usa.philips.com/healthcare/solutions/customer-service-solutions team for additional guidance. Please see the Philips product security website https://www.philips.com/productsecurity for the latest security information for Philips products.
Workaround
As an interim mitigation to this vulnerability, Philips recommends the following: * The IBE transaction logs are only accessible with administrative privileges. An additional account can be created on the IBE system with limited privileges, for service engineers. * Reduce log retention to an acceptable timeframe that allows for recovery activities. Users with questions regarding their specific Philips IntelliBridge Enterprise installation should contact their local Philips service support team, or regional service support. Users can see the Philips advisory http://www.philips.com/productsecurity for more details, and contact the Philips Customer Service Solutions https://www.usa.philips.com/healthcare/solutions/customer-service-solutions team for additional guidance. Please see the Philips product security website https://www.philips.com/productsecurity for the latest security information for Philips products.
Wed, 04 Jun 2025 22:15:00 +0000
| Type | Values Removed | Values Added |
|---|---|---|
| Title | Philips IntelliBridge Enterprise IBE Insertion of Sensitive Information into Log File | |
| References |
| |
| Metrics |
cvssV3_1
|
cvssV3_1
|
Status: PUBLISHED
Assigner: icscert
Published:
Updated: 2025-06-04T22:02:51.755Z
Reserved: 2020-04-21T00:00:00
Link: CVE-2020-12023
No data.
Status : Modified
Published: 2020-06-11T20:15:10.960
Modified: 2025-06-04T22:15:22.557
Link: CVE-2020-12023
No data.
OpenCVE Enrichment
No data.
EUVD