Metrics
Affected Vendors & Products
| Source | ID | Title |
|---|---|---|
EUVD |
EUVD-2020-17857 | An improper verification of the cryptographic signature of firmware updates of the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers to generate valid firmware updates with arbitrary content that can be used to tamper with devices. |
Solution
B. Braun recommends applying updates: SpaceCom: Version U62 or later (United States), L82 or later (outside the United States) Battery Pack SP with Wi-Fi: Version U62 or later (United States), L82 or later (outside the United States) Data module compactplus: Version A12 or later Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html
Workaround
As a general security measure, B. Braun recommends protecting the network with appropriate mechanisms: Ensure the devices are not accessible directly from the Internet. Use a firewall and isolate the medical devices from the business network. Please contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html
Wed, 16 Apr 2025 17:15:00 +0000
| Type | Values Removed | Values Added |
|---|---|---|
| Metrics |
ssvc
|
Status: PUBLISHED
Assigner: icscert
Published:
Updated: 2025-04-16T16:29:44.744Z
Reserved: 2020-09-04T00:00:00.000Z
Link: CVE-2020-25166
Updated: 2024-08-04T15:26:10.204Z
Status : Modified
Published: 2022-04-14T21:15:08.297
Modified: 2024-11-21T05:17:31.640
Link: CVE-2020-25166
No data.
OpenCVE Enrichment
No data.
EUVD