Show plain JSON{"containers": {"cna": {"affected": [{"product": "ZOOM LATITUDE", "vendor": "Boston Scientific", "versions": [{"status": "affected", "version": "Model 3120"}]}], "credits": [{"lang": "en", "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices and reported them to Boston Scientific."}], "datePublic": "2021-09-30T00:00:00", "descriptions": [{"lang": "en", "value": "The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB."}], "metrics": [{"cvssV3_1": {"attackComplexity": "LOW", "attackVector": "PHYSICAL", "availabilityImpact": "LOW", "baseScore": 6.5, "baseSeverity": "MEDIUM", "confidentialityImpact": "LOW", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "CHANGED", "userInteraction": "REQUIRED", "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L", "version": "3.1"}}], "problemTypes": [{"descriptions": [{"cweId": "CWE-353", "description": "CWE-353 Missing Support for Integrity Check", "lang": "en", "type": "CWE"}]}], "providerMetadata": {"dateUpdated": "2021-10-04T17:35:21", "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert"}, "references": [{"tags": ["x_refsource_MISC"], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"}], "source": {"advisory": "ICSMA-21-273-01", "defect": ["CWE-353"], "discovery": "EXTERNAL"}, "title": "Missing Support Integrity Check for Boston Scientific Zoom Latitude", "workarounds": [{"lang": "en", "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."}], "x_generator": {"engine": "Vulnogram 0.0.9"}, "x_legacyV4Record": {"CVE_data_meta": {"ASSIGNER": "ics-cert@hq.dhs.gov", "DATE_PUBLIC": "2021-09-30T21:02:00.000Z", "ID": "CVE-2021-38396", "STATE": "PUBLIC", "TITLE": "Missing Support Integrity Check for Boston Scientific Zoom Latitude"}, "affects": {"vendor": {"vendor_data": [{"product": {"product_data": [{"product_name": "ZOOM LATITUDE", "version": {"version_data": [{"version_value": "Model 3120"}]}}]}, "vendor_name": "Boston Scientific"}]}}, "credit": [{"lang": "eng", "value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices and reported them to Boston Scientific."}], "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": {"description_data": [{"lang": "eng", "value": "The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB."}]}, "generator": {"engine": "Vulnogram 0.0.9"}, "impact": {"cvss": {"attackComplexity": "LOW", "attackVector": "PHYSICAL", "availabilityImpact": "LOW", "baseScore": 6.5, "baseSeverity": "MEDIUM", "confidentialityImpact": "LOW", "integrityImpact": "HIGH", "privilegesRequired": "NONE", "scope": "CHANGED", "userInteraction": "REQUIRED", "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L", "version": "3.1"}}, "problemtype": {"problemtype_data": [{"description": [{"lang": "eng", "value": "CWE-353 Missing Support for Integrity Check"}]}]}, "references": {"reference_data": [{"name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01", "refsource": "MISC", "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"}]}, "source": {"advisory": "ICSMA-21-273-01", "defect": ["CWE-353"], "discovery": "EXTERNAL"}, "work_around": [{"lang": "en", "value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."}]}}, "adp": [{"providerMetadata": {"orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE", "dateUpdated": "2024-08-04T01:37:16.545Z"}, "title": "CVE Program Container", "references": [{"tags": ["x_refsource_MISC", "x_transferred"], "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"}]}]}, "cveMetadata": {"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2021-38396", "datePublished": "2021-10-04T17:35:21.072612Z", "dateReserved": "2021-08-10T00:00:00", "dateUpdated": "2024-09-16T20:46:57.503Z", "state": "PUBLISHED"}, "dataType": "CVE_RECORD", "dataVersion": "5.1"}