A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.



Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution:
Customers should ensure the following:
• The network is secure, and access follows best practices.
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                     
Exploit Status:


Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
Advisories

No advisories yet.

Fixes

Solution

Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.


Workaround

If the network is not considered secure, please remove the analyzer from the network.

References
History

Wed, 17 Dec 2025 13:00:00 +0000

Type Values Removed Values Added
Description A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
Title Remote Code Execution Vulnerability in Radiometer Products
Weaknesses CWE-287
References
Metrics cvssV3_1

{'score': 7.2, 'vector': 'CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H'}


Projects

Sign in to view the affected projects.

cve-icon MITRE

Status: PUBLISHED

Assigner: Radiometer

Published:

Updated: 2025-12-17T14:41:13.252Z

Reserved: 2025-12-05T10:50:03.683Z

Link: CVE-2025-14097

cve-icon Vulnrichment

No data.

cve-icon NVD

Status : Received

Published: 2025-12-17T13:15:58.050

Modified: 2025-12-17T13:15:58.050

Link: CVE-2025-14097

cve-icon Redhat

No data.

cve-icon OpenCVE Enrichment

No data.

Weaknesses