Impact
A path traversal flaw in the SCPI FILESystem command interface of Tektronix 3 Series MDO oscilloscopes allows an attacker to craft file paths that bypass directory restrictions, thereby reading or modifying arbitrary firmware and configuration files on the device. This vulnerability is classified as CWE-22 and could enable further compromise if critical system files are altered.
Affected Systems
Tektronix 3 Series MDO oscilloscopes with firmware versions earlier than 1.21.1 are affected.
Risk and Exploitability
Based on the description, it is inferred that the attack vector is remote exploitation via the SCPI protocol, requiring no authentication. The CVSS score of 8.1 indicates high severity, and the remote nature of the flaw suggests a broad attack surface. While the EPSS score is unavailable and the vulnerability is not listed in CISA KEV, the high CVSS makes it a serious risk that should be addressed promptly.