Description
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Published: 2019-03-25
Score: 9.8 Critical
EPSS: < 1% Very Low
KEV: No
Impact: n/a
Action: n/a
AI Analysis

No analysis available yet.

Remediation

No remediation available yet.

Tracking

Sign in to view the affected projects.

Advisories
Source ID Title
EUVD EUVD EUVD-2015-3987 Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
History

No history.

Subscriptions

Pifzer Plum A\+3 Infusion System Plum A\+3 Infusion System Firmware Plum A\+ Infusion System Plum A\+ Infusion System Firmware Symbiq Infusion System Symbiq Infusion System Firmware
cve-icon MITRE

Status: PUBLISHED

Assigner: icscert

Published:

Updated: 2024-08-06T06:04:02.927Z

Reserved: 2015-05-12T00:00:00.000Z

Link: CVE-2015-3956

cve-icon Vulnrichment

No data.

cve-icon NVD

Status : Modified

Published: 2019-03-25T18:29:00.323

Modified: 2024-11-21T02:30:08.517

Link: CVE-2015-3956

cve-icon Redhat

No data.

cve-icon OpenCVE Enrichment

No data.

Weaknesses